## Abstract

Introduction: This study aims to demonstrate the importance of the finite element method (FEM) in total knee arthroplasty (TKA) to minimize the necessity for revision. Materials and Methods: In our study, the TKA system implanted in femoral and tibial bones has been numerically analyzed by three-dimensional FEM. The results and the TKA model have been transferred to 3D SolidWorks 3D software. This study's application aspect focuses on the stress distribution in the interaction between the distal bone portions of the knee joint and the TKA. Our study presents the load distribution on the implants and the demonstration data applied in the normal surgical procedure. The load distribution has also been analyzed, and the degree of valgus was reduced by 3 degrees. This situation helped us to predict early polyethylene corrosion and failure in the future. Results: When the valgus angle was reduced by 3°, the load distribution on the implants increased in the medial compartment, especially in the polyethylene insert. This led to the prediction of early polyethylene wear and failure in the future. Conclusion: FEM emphasizes the importance of biomechanical load distribution during TKA. It is also a good method to reduce the incidence of TKA revision and increase patient satisfaction after TKA.

## Keywords

Finite element analysis; total knee arthroplasty; polyethylene wear knee

## How to cite

Azarabadi JM. Evaluation of Total Knee Arthroplasty Surgery with Finite Element Analysis. Ulutas Med J. 2024;10(2):39-43. doi:10.5455/umj.20240305070048

## References

1. Çeliktaş M, SARPEL DY. Enfekte total diz protezlerinde iki aşamalı revizyon yapilan hastalarda orta ve uzun dönem sonuçlar. Çukurova Üniversitesi. 2007.
2. Khan M, Osman K, Green G, Haddad F. The epidemiology of failure in total knee arthroplasty: avoiding your next revision. The bone & joint journal. 2016;98(1_Supple_A):105-12. https://doi.org/10.1302/0301-620x.98b1.36293
3. Bülbül AM, Batmaz AG. Diz artroplastisinin revizyon komplikasyonları. TOTBID Dergisi. 2015;14(2):140-4. https://doi.org/10.14292/totbid.dergisi.2015.20
4. Popescu R, Haritinian EG, Cristea S. Relevance of finite element in total knee arthroplasty-Literature Review. Chirurgia (Bucharest, Romania: 1990). 2019;114(4):437-42. https://doi.org/10.21614/chirurgia.114.4.437
5. Au AG, Liggins AB, Raso VJ, Amirfazli A. A parametric analysis of fixation post shape in tibial knee prostheses. Medical Engineering & Physics. 2005;27(2):123-34. https://doi.org/10.1016/j.medengphy.2004.09.010
6. Latypova A, Pioletti D, Terrier A. Importance of trabecular anisotropy in finite element predictions of patellar strain after total knee arthroplasty. Medical Engineering & Physics. 2017;39:102-5. https://doi.org/10.1016/j.medengphy.2016.10.003
7. Sahu NK, Kaviti AK. A review of use FEM techniques in modeling of human knee joint. Journal of Biomimetics, Biomaterials and Biomedical Engineering. 2016;28:14-25. https://doi.org/10.4028/www.scientific.net/jbbbe.28.14
8. Hutter EE, Granger JF, Beal MD, Siston RA. Is there a gold standard for TKA tibial component rotational alignment? Clinical Orthopaedics and Related Research®. 2013;471(5):1646-53. https://doi.org/10.1007/s11999-013-2822-0
9. Henak CR, Anderson AE, Weiss JA. Subject-specific analysis of joint contact mechanics: application to the study of osteoarthritis and surgical planning. Journal of biomechanical engineering. 2013;135(2). https://doi.org/10.1115/1.4023386
10. Li J, Hua X, Jin Z, Fisher J, Wilcox RK. Influence of clearance on the time-dependent performance of the hip following hemiarthroplasty: A finite element study with biphasic acetabular cartilage properties. Medical engineering & physics. 2014;36(11):1449-54. https://doi.org/10.1016/j.medengphy.2014.05.014
11. Boccaccio A, Ballini A, Pappalettere C, Tullo D, Cantore S, Desiate A. Finite element method (FEM), mechano-biology and biomimetic scaffolds in bone tissue engineering. International journal of biological sciences. 2011;7(1):112-32. https://doi.org/10.7150/ijbs.7.112
12. Mahajan S, Patil R. Optimizing dental implant model by comparison of three dimensional finite element. Int J Innov Res Sci Eng Technol. 2015;4:11223-31.
13. Arab AZE-A, Merdji A, Benaissa A, Roy S, Bouiadjra B-AB, Layadi K, et al. Finite-element analysis of a lateral femoro-tibial impact on the total knee arthroplasty. Computer methods and programs in biomedicine. 2020;192:105446. https://doi.org/10.1016/j.cmpb.2020.105446
14. Mou Z, Dong W, Zhang Z, Wang A, Hu G, Wang B, et al. Optimization of parameters for femoral component implantation during TKA using finite element analysis and orthogonal array testing. Journal of Orthopaedic Surgery and Research. 2018;13(1):1-12. https://doi.org/10.1186/s13018-018-0891-1
15. Azam N, Daud R, Ayu HM, Ramli J, Hassan M, Shah A, et al., editors. The effect of knee flexion angle on contact stress of total knee arthroplasty. MATEC Web of Conferences; 2018: EDP Sciences. https://doi.org/10.1051/matecconf/201822503009
16. Kang K-T, Son J, Kwon O-R, Koh Y-G. Malpositioning of prosthesis: patient-specific total knee arthroplasty versus standard off-the-shelf total knee arthroplasty. Journal of the American Academy of Orthopaedic Surgeons Global Research & Reviews. 2017;1(4). https://doi.org/10.5435/jaaosglobal-d-17-00020
17. Moewis P, Checa S, Kutzner I, Hommel H, Duda GN. Physiological joint line total knee arthroplasty designs are especially sensitive to rotational placement–A finite element analysis. PLoS One. 2018;13(2):e0192225. https://doi.org/10.1371/journal.pone.0192225
18. Healy WL, Della Valle CJ, Iorio R, Berend KR, Cushner FD, Dalury DF, et al. Complications of total knee arthroplasty: standardized list and definitions of the Knee Society. Clinical Orthopaedics and Related Research®. 2013;471(1):215-20. https://doi.org/10.1007/s11999-012-2489-y
19. Çalbıyık M. Clinical outcome of total knee arthroplasty performed using patient-specific cutting guides. Medical Science Monitor: International Medical Journal of Experimental and Clinical Research. 2017;23:6168. https://doi.org/10.12659/msm.908213
20. Robertsson O, Dunbar M, Pehrsson T, Knutson K, Lidgren L. Patient satisfaction after knee arthroplasty: a report on 27,372 knees operated on between 1981 and 1995 in Sweden. Acta Orthopaedica Scandinavica. 2000;71(3):262-7. 43 77 Total Knee Arthroplasty with Finite Element Analysis Javid Mohammadzadeh Azarabadi https://doi.org/10.1080/000164700317411852
21. Abolghasemian M, Samiezadeh S, Sternheim A, Bougherara H, Barnes CL, Backstein DJ. Effect of patellar thickness on knee flexion in total knee arthroplasty: a biomechanical and experimental study. The Journal of arthroplasty. 2014;29(1):80-4. https://doi.org/10.1016/j.arth.2013.04.026
22. Norman TL, Hutchison JD, Gardner MR, Blaha MD J. Knee loading due to varus and external rotation in gait supports medial compartment wear in total knee arthroplasty. Journal of Orthopedics and Rheumatism. 2017;1(1):8.
23. Browne C, Hermida JC, Bergula A, Colwell Jr CW, D'Lima DD. Patellofemoral forces after total knee arthroplasty: effect of extensor moment arm. The Knee. 2005;12(2):81-8. https://doi.org/10.1016/j.knee.2004.05.006
24. Ostermeier S, Stukenborg-Colsman C. Quadriceps force after TKA with femoral single radius: an in vitro study. Acta orthopaedica. 2011;82(3):339-43. https://doi.org/10.3109/17453674.2011.574564
25. Gallo J, Goodman SB, Konttinen YT, Wimmer MA, Holinka M. Osteolysis around total knee arthroplasty: a review of pathogenetic mechanisms. Acta biomaterialia. 2013;9(9):8046-58. https://doi.org/10.1016/j.actbio.2013.05.005
26. Sathasivam S, Walker PS, Campbell PA, Rayner K. The effect of contact area on wear in relation to fixed bearing and mobile bearing knee replacements. Journal of Biomedical Materials Research: An Official Journal of The Society for Biomaterials, The Japanese Society for Biomaterials, and The Australian Society for Biomaterials and the Korean Society for Biomaterials. 2001;58(3):282-90. https://doi.org/10.1002/1097-4636(2001)58:3<282::aid-jbm1018>3.3.co;2-j
27. Tanska P, Mononen ME, Korhonen RK. A multi-scale finite element model for investigation of chondrocyte mechanics in normal and medial meniscectomy human knee joint during walking. Journal of biomechanics. 2015;48(8):1397-406. https://doi.org/10.1016/j.jbiomech.2015.02.043